Abstract

A cryogenic sampler for aniline vapors was evaluated by collecting samples of 10 ppm aniline in air and directly analyzing the collected aniline by liquid chromatography with electrochemical detection. The cryogenic sampler achieved nearly 100% collection and recovery efficiency, with a detection limit of ≈ 4 parts-per-billion (10 9) aniline in air for a sample collected at 10 cm 3/min for 40 s in a 200-μl stainless-steel loop at — 10°C. About 5 min was necessary to dissolve the trapped aniline into 200 μl of a mobile phase desorbing solution at 20°C.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.