Abstract

When a dye solution used to simulate patient material was either injected into high-speed dental handpiece (drill) waterlines or applied to the equipment externally, internal air turbine chambers became contaminated. These chambers served as a reservoir of the material, which was slowly dislodged by air expelled during subsequent handpiece operation and which was diluted by water spray used for cooling the drilling surface. Considering the fact that patient materials could reside in internal parts of the equipment that are not usually disinfected and that the material may be subsequently sprayed into cuts and abrasions in the oral cavity, the common approach to reprocessing handpieces (external wiping in combination with flushing) may pose unacceptably high risks to those individuals treated soon after infected patients. Therefore, unless reliable data on cross-infection frequencies are obtained and prove it unnecessary, heat-treating high-speed handpieces between each patient should be considered an essential component of standard procedures whenever universal precautions are practiced in dentistry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.