Abstract

High breast tissue density has been shown to be one of the strongest predictors of breast cancer risk in women; though currently there are no calibrated methods to ensure that clinical measures of density are accurate. This inability to make accurate measurements impedes the use of breast tissue density as a cancer risk indicator. An accurate and precise method to measure the volumetric density of breast tissue in a clinical environment has been developed without modifying existing methods. The single x-ray absorptiometry (SXA) approach is validated against known breast fibroglandular density test phantoms and the accuracy and precision of the measure is monitored longitudinally. With cross-calibration and longitudinal quality control using the same test phantoms we show that SXA density measures are comparable between seven Hologic Selenia Full-Field Digital Mammography systems. In the future, we plan to address whether SXA measures can be generalized across different makes and models.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.