Abstract

Background and purposeDelta4 (ScandiDos AB, Uppsala, Sweden) and ArcCHECK with 3DVH software (Sun Nuclear Corp., Melbourne, FL, USA) are commercial quasi-three-dimensional diode dosimetry arrays capable of volumetric measurement-guided dose reconstruction. A method to reconstruct dose for non-coplanar VMAT beams with 3DVH is described. The Delta4 3D dose reconstruction on its own phantom for VMAT delivery has not been thoroughly evaluated previously, and we do so by comparison with 3DVH. Materials and methodsReconstructed volumetric doses for VMAT plans delivered with different table angles were compared between the Delta4 and 3DVH using gamma analysis. ResultsThe average γ (2% local dose-error normalization/2mm) passing rate comparing the directly measured Delta4 diode dose with 3DVH was 98.2±1.6% (1SD). The average passing rate for the full volumetric comparison of the reconstructed doses on a homogeneous cylindrical phantom was 95.6±1.5%. No dependence on the table angle was observed. ConclusionsModified 3DVH algorithm is capable of 3D VMAT dose reconstruction on an arbitrary volume for the full range of table angles. Our comparison results between different dosimeters make a compelling case for the use of electronic arrays with high-resolution 3D dose reconstruction as primary means of evaluating spatial dose distributions during IMRT/VMAT verification.