Abstract

Development of a compact fluorescence-based detection system for use in a micro-analytical system, such as a point-of-care diagnostic system, often requires a multi-channel microfluidic chip system. Since the materials used for microfluidic chips usually are transparent in the visible region and have a refractive indices higher than that of air or the surrounding environment, the fluorescence emission and scattered excitation light can propagate through the chip. We observed that such propagation can cause cross-talk between adjacent channels, and may become the major source of noise in the system and/or photo bleach the fluorescent samples in the adjacent channels, particularly for the small distances between the channels found in microfluidic chips, usually in order of several micro m. We monitored this cross-talk using fluorescein as a fluorescent sample and Mylar sheeting as a microfluidic chip material. We then discuss how this cross-talk can be avoided using a simple, inexpensive and effective method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.