Abstract

Background: The AmbuAuraGain is a new single-use supraglottic airway device with gastric channel designed to facilitate intubation. The aim of the study was to assess the anatomic position and the performance of the AuraGain in fresh cadavers compared to that of the Intersurgical i-gel and LMA Supreme New Cuff. Methods: The 3 devices were inserted in random order in 7 fresh cadavers without difficult airway criteria. The assessed items were: Insertion time, number of attempts and ease of insertion, airway seal pressure, ease of gastric tube insertion, endoscopic view of vocal cords, efficacy of guided tracheal intubation through the AuraGain and i-gel, and anatomic fit by lateral X-ray and neck dissections. Results: All devices were successfully inserted within 3 attempts, except for one case of the LMA Supreme. Adjusting manoeuvres were often required to accomplish correct insertion. A 16 G gastric tube was easily advanced through all AuraGain and LMA Supreme devices. Fiberoptic tracheal intubation was effectively achieved through all AuraGain and i-gel devices in less than 60 s. Lateral X-ray and neck dissections confirmed optimal alignment of all devices with the respiratory and digestive tracts. Conclusions: Insertion of the new AmbuAuraGain required adjusting manoeuvres in some cases, as observed with the other two devices, and achieved similar airway seal pressures. Passage of a large bore gastric tube was as fast as with the LMA Supreme and ease of guided intubation was similar to that of the i-gel.

Highlights

  • The use of second generation supraglottic airway devices (SAD) is increasing exponentially since their introduction [1]

  • The airway conduit does not permit a direct passage of a tracheal tube (TT) but fibreoptic guided intubation is possible with the aid of tube introducers

  • Dissection of two necks showed the correct position of the bowl around the glottis inlet with all three brands of SGAs assessed (Figure 5). This pre-clinical evaluation in fresh cadavers shows that the performance of the AuraGain is in line with that of the i-gel or LMA Supreme with respect to ease of insertion, anatomic position, efficacy of gastric access and capability to serve as an intubating conduit

Read more

Summary

Introduction

The use of second generation supraglottic airway devices (SAD) is increasing exponentially since their introduction [1]. The AmbuAuraGain is a new single-use supraglottic airway device with gastric channel designed to facilitate intubation. The aim of the study was to assess the anatomic position and the performance of the AuraGain in fresh cadavers compared to that of the Intersurgical i-gel and LMA Supreme New Cuff. The assessed items were: Insertion time, number of attempts and ease of insertion, airway seal pressure, ease of gastric tube insertion, endoscopic view of vocal cords, efficacy of guided tracheal intubation through the AuraGain and i-gel, and anatomic fit by lateral X-ray and neck dissections. Conclusions: Insertion of the new AmbuAuraGain required adjusting manoeuvres in some cases, as observed with the other two devices, and achieved similar airway seal pressures. Passage of a large bore gastric tube was as fast as with the LMA Supreme and ease of guided intubation was similar to that of the i-gel

Objectives
Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.