Abstract

This article presents a novel real-time meta-material (MM) sensor based on a non-invasive method that operates in microwave frequency ranges at 8.524 GHz to measure blood glucose levels with quality factor 184 is designed and fabricated. A cross enclosed between two square shapes produces a strong interaction between glucose samples and electromagnetic waves. In this study, 5 were tested noninvasively using the proposed glucose resonant sensor with a range of glucose-level changes from 50 to 130 mg/dL. For this range of glucose-level changes, the frequency detection resolution is 5.06 MHz/(mg/dL), respectively. Despite simulations, fabrication procedures (F.P.) have been carried out for more precise result verification. For the purpose of qualitative analysis, the proposed MM sensor is considered a viable candidate for determining glucose levels.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.