Abstract

The primary magnetic field fluctuation thermometer (pMFFT) is a SQUID-based noise thermometer for temperatures below 1 K, which complies with metrological requirements. It combines two signal channels in order to apply the cross-correlation technique, but it requires statistically independent noise signals for proper operation. In order to check the limit of the cross-correlation readout, we have performed zero measurements in the millikelvin range in a setup that is identical to the pMFFT, except for the removed temperature sensor. We examined the influence of different parameters such as SQUID working point or flux-lock loop parameters on the minimum cross-correlation signal down to 24 mK and below 100 kHz. Depending on the configuration, typical minimum SQUID-referred cross-power spectral densities of 1.5 × 10−15 /Hz or even smaller values were observed. For the pMFFT, considering its thermal noise spectrum, these flux densities correspond to a device noise temperature of ≤2.5 µK, thereby ensuring a negligible uncertainty contribution at the lower end of the PLTS-2000 (0.9 mK).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.