Abstract

The present work gives an insight into some key measurement and signal processing considerations in terahertz pulsed imaging (TPI). TPI is increasingly used for the measurement of the spatial variation of coating thickness on coated solid dosage forms. The potential of TPI for the assessment of coating thickness distributions and the use in process development is described in recent literature. However, some critical factors need to be taken into account when working with this technique. These are (1) the signal processing of the raw data, (2) the influence of the composition of the sample matrix on the TPI signals and subsequent coating analysis, (3) signal distortions that can occur at tablet edges or areas with defects, and (4) the refractive index as a key parameter in the quantification of layer thickness. In this paper, we will highlight to what extent these factors impact on the qualitative and quantitative analysis of TPI data and how artifacts and misinterpretation of data can be avoided to ensure fully quantitative and robust measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.