BackgroundTo evaluate and establish a digital workflow for the custom designing and 3D printing of mouth opening tongue-depressing (MOTD) stents for patients receiving radiotherapy for head and neck cancer.MethodsWe retrospectively identified 3 patients who received radiation therapy (RT) for primary head and neck cancers with MOTD stents. We compared two methods for obtaining the digital impressions of patients’ teeth. The first method involved segmentation from computed tomography (CT) scans, as previously established by our group, and the second method used 3D scanning of the patients’ articulated stone models that were made during the conventional stent fabrication process. Three independent observers repeated the process to obtain digital impressions which provided data to design customized MOTD stents. For each method, we evaluated the time efficiency, dice similarity coefficient (DSC) for reproducibility, and the 3D printed stents’ accuracy. For the 3D scanning method, we evaluated the registration process using manual and automatic approaches.ResultsFor all patients, the 3D scanning method demonstrated a significant advantage over the CT scanning method in terms of time efficiency with over 60% reduction in time consumed (p < 0.0001) and reproducibility with significantly higher DSC (p < 0.001). The printed stents were tested over the articulated dental stone models, and the trueness of fit and accuracy of dental anatomy was found to be significantly better for MOTD stents made using the 3D scanning method. The automated registration showed higher accuracy with errors < 0.001 mm compared to manual registration.ConclusionsWe developed an efficient workflow for custom designing and 3D-printing MOTD radiation stents. This workflow represents a considerable improvement over the CT-derived segmentation method. The application of this rapid and efficient digital workflow into radiation oncology practices can expand the use of these toxicity sparing devices to practices that do not currently have the support to make them.

Highlights

  • To evaluate and establish a digital workflow for the custom designing and 3D printing of mouth opening tongue-depressing (MOTD) stents for patients receiving radiotherapy for head and neck cancer

  • This method is limited by availability of high quality scans, absence of dental artifacts, subjectivity in delineation of the dental anatomy, and bite registration inaccuracy, as the mandible is rotated around an anatomical average rather than a patient specific axis

  • Time and reproducibility (CT vs. 3D-scanning) Results showed a statistically significant difference in the time consumed for acquiring the digital dental impressions between the two methods

Read more

Summary

Introduction

To evaluate and establish a digital workflow for the custom designing and 3D printing of mouth opening tongue-depressing (MOTD) stents for patients receiving radiotherapy for head and neck cancer. Therapeutic modalities and devices have been developed to tackle RIOM, and one of the most utilized devices is the oral radiation stent. We demonstrated the feasibility of utilizing computer-aided-designing (CAD) and 3D-printing to create oral stents using routine diagnostic CT imaging studies [13]. This method is limited by availability of high quality scans, absence of dental artifacts, subjectivity in delineation of the dental anatomy, and bite registration inaccuracy, as the mandible is rotated around an anatomical average rather than a patient specific axis. We investigated the utility and limitations in acquiring accurate and reproducible teeth impressions, and subsequent performance of 3D printed customized MOTD stents

Methods
Results
Discussion
Conclusion