Abstract
Inpatients (n = 57) on long term prophylaxis with 2 oral phenytoin preparations were followed up via monthly checks of serum drug concentrations. Duplicate serum aliquots were submitted to 2 laboratories, and covariance analysis was used to estimate laboratory error. The laboratory-associated variance of examinations using the 'EMIT' or 'TDX' systems was 7.1 to 10 (mg/L)2, while that of chromatographic assays was 37.1 (mg/L)2. Laboratory errors were distributed in a non-normal fashion. The ratio of the maximum rate of metabolism (Vmax) and the Michaelis-Menten constant (Km) tended to remain constant between individuals. A linear analytical model showed little more residual variance than one based on Michaelis-Menten pharmacokinetics.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.