Abstract

The concept of interfacing a large-size column for capillary electrophoresis (CE) to electrospray ionization mass spectrometry (ESI-MS) for robust and automatic CE–MS operation is reported. Both standard ionspray interface and microionspray interface have been modified to operate in a sheath flow pattern to overcome the common stability problem in CE–MS coupling. To make the interface sensitive, a step-down stainless steel tube with smaller inner diameter and tapered tip was incorporated onto a larger tube embracing the CE column via cold soldering. The devices were evaluated for quantitative analysis of nucleotides at femtomole level and stable analytical performance in peptide profiling.