Abstract

Most applications of superconducting quantum interference devices (SQUID’s) involve use of a coupling coil. There are two classes of such applications: magnetometry (magnetic field, field gradient) and voltage or current measurements. We have analyzed the factors which determine the ultimate sensitivity, and present a guide for optimization of signal/noise in each of the two cases. It is found that a single figure of merit, involving both the equivalent input noise of the SQUID and the effectiveness of the coupling to the coil, is applicable to either type of measurement. We propose that this figure of merit be used in evaluating the performance of new SQUID designs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.