Abstract

Despite the recent success of coupling anion exchange chromatography with native mass spectrometry (AEX-MS) to study anionic proteins, the utility of AEX-MS methods in therapeutic monoclonal antibody (mAb) characterization has been limited. In this work, we developed and optimized a salt gradient-based AEX-MS method and explored its utility in charge variant analysis of therapeutic mAbs. We demonstrated that, although the developed AEX-MS method is less useful for IgG1 molecules that have higher isoelectric points (pIs), it is an attractive alternative for charge variant analysis of IgG4 molecules. By elevating the column temperature and lowering the mAb pI through PNGase F-mediated deglycosylation, the chromatographical resolution from AEX separation can be significantly improved. We also demonstrated that, after PNGase F and IdeS digestion, the AEX-MS method exhibited excellent resolving power for multiple attributes in the IgG4 Fc region, including unprocessed C-terminal Lys, N-glycosylation occupancy, and several conserved Fc deamidations, making it ideally suited for multiple attribute monitoring (MAM). Through fractionation and peptide mapping analysis, we also demonstrated that the developed AEX-MS method can provide site-specific and isoform-resolved separation of Fc deamidation products, allowing rapid and artifact-free quantitation of these modifications without performing bottom-up analysis.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call