Abstract
Synchrotron light sources are low-duty-cycle pulsed X-ray sources, a fact that is often neglected in estimating the count-rate capabilities of photon-counting detectors in synchrotron-based experiments. In this paper, we demonstrate the effect that this has on the pileup statistics of drift detectors. We derive expressions for the cases of continuous and pulsed X-ray sources. We consider a pulsed source with period that is either much less than the shaper support time or much less than the average drift time. We also consider a pulsed source with a period that is long or comparable to both the shaper support and the drift time. These conditions correspond to normal and reduced bunch fill patterns of synchrotrons.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.