Abstract

Measuring the plasma corticotropin (ACTH) concentration is an important step in determining the underlying cause in patients with hypercortisolemia. Interfering substances in immunoassays can lead to erroneous results impacting clinical management. We describe a case series of 12 patients, the majority of whom were being investigated for possible Cushing's syndrome and in whom inconsistencies between the clinical picture and biochemical testing raised concerns of assay interference. ACTH assay interference resulted in falsely elevated ACTH concentrations using the Siemens Immulite assay and consequently led to additional unnecessary testing. Communication between physician and laboratory as well as appropriate investigation (including sample dilution, use of blocking antibodies and testing on an alternate platform) resulted in assay interference identification. Recognition of biochemical results which are clinically discrepant remains an essential step in patient assessment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.