Abstract

AbstractIn this study, experimental Ti–6Al–1V–3Fe, Ti–4.5Al–1V–3Fe, and Ti–3Fe alloys, as well as commercial Ti–6Al–4V alloy that were scaled up utilizing vacuum induction melting technology, were assessed for corrosion performance in simulated body fluids. The selected simulated body fluids were 0.9 wt% NaCl solution and Hanks balanced salt solution (HBSS). Open circuit potential and linear polarization scans were performed to understand the corrosion performance of the alloys. The surface of the alloys was examined before and after exposure to corrosive solutions using scanning electron microscopy. The results show that all the alloys exhibit good corrosion performance in simulated body fluids. The corrosion rates were less than 0.5 mm/year. Owing to higher corrosion potential and lower corrosion rate, Ti–6Al–1V–3Fe and Ti–4.5Al–1V–3Fe had the best resistance to corrosion in 0.9 wt% NaCl and HBSS, respectively. All the alloys consist of a fully lamellar structure with α and β phases. There was no evidence of severe deterioration on the exposed surface of alloys in the simulated body fluids.