Abstract

The anodic oxidation/TiO2 composite film was prepared on the surface of AZ31B Mg alloy by DC magnetron sputtering. The corrosion resistance and blood compatibility of the film were systematically studied by electrochemical, dynamic clotting time and platelet adhesion test. The results shows that the corrosion current of AZ31B Mg alloy was 6.409×10-8A/cm2 after anodic oxidation treatment, which has decreased 4 orders of magnitude compared to the untreated samples and the corrosion resistance is improved greatly. The clotting time of anodic oxidation/TiO2 film is about 53 min, which has increased 1.3 times compared to anodic oxidation film (40min). Platelets adhesion to anodic oxidation/TiO2 film are less than the one adhesion to anodic oxidation film, and there are no pseudopodia and aggregation, which indicate that the blood compatibility of anodic oxidation/TiO2 film is better than anodic oxidation film.