Abstract

Surface treatment of medical CoCr alloys L605 by nitrogen plasma immersion ion implantation (PIII) leads to the formation of a hard and wear resistant surface layer, consisting of nitrogen in solid solution. However, a detailed investigation of the corrosion properties by potentiodynamic polarization and electrochemical impedance spectroscopy shows that even at processing temperatures of 350 °C, where no CrN precipitates are observed, no complete passivation of the surface by formation of a protective Cr 2O 3 layer is possible leading to enhanced corrosion rates further increasing with increasing PIII processing temperature. It is postulated that the enhanced affinity of chromium for nitrogen leads to a reduced mobility inside the alloy, thus prohibiting a timely surface passivation. Nevertheless, a surface modification where a moderate decrease in corrosion resistance coupled with a significant reduction in generated wear particles should be feasible for biomedical applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.