Abstract
With CP Ti sample as control, the electrochemical measurements were used to investigate the corrosion resistance of low modulus Ti16Nb shape memory alloy in Hank's solution with pH value 7.4. The OCP curves show that the passive film on the Ti16Nb alloy surface is quite stable. The tafel plots and anodic polarization curves prove that the Ti16Nb alloy has excellent anti-corrosion properties in Hank's solution, which is even better than that of CP Ti. The XPS analysis reveals that the composition of the oxide film on Ti16Nb alloy is mainly TiO 2 and Nb 2O 5. The indirect cytotoxicity results prove that Ti16Nb alloy has an excellent biocompatibility as CP Ti with high proliferation rate approaching that of the negative control group.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.