Abstract

Two types of HL hexed abutments of a Steri-Oss system, gold/plastic coping and gold coping, were compared in terms of corrosion behavior. The anodic polarization behavior and the galvanic corrosion between abutments and Type III gold alloys, before and after casting, were analyzed. In addition, the crevice corrosion of the casting samples was analyzed with cyclic potentiodynamic polarization tests using the |Er-Ecorr| value and scanning electron microscopy. Before casting, gold/plastic coping and gold coping were shown to have similar corrosion patterns in the anodic polarization test. Type III casting gold alloy was shown to have a lower higher than that of gold coping, but the passive region for the gold/plastic coping was smaller than that of gold coping. The contact current density between the cast gold alloys and gold/plastic before casting was higher than that between gold coping and cast gold alloy. The contact current density of the samples after casting was shown to be similar to that before casting. The crevice corrosion resistance of cast samples using gold coping was lower than that of cast samples using gold/plastic coping, and severe corrosion was observed by SEM at the abutment-casting gold alloy interface.