Poor crystallinity remains a major problem affecting the availability and cost of CdZnTe (CZT) detectors. Point defects are responsible for small gradual charge loss and correlated with the electron clouds' drift times, which allows electronic correction of the output signals to achieve high spectral-resolution even with large-volume CZT detectors. In contrast, extended defects causes significant charge losses, which typically are uncorrelated, and, thus, result in much greater fluctuations of the output signals that cannot be corrected. Although extended defects do not affect all the interaction events, their fraction rapidly increases with the crystal's thickness and volume. In this paper, we summarize our recent results from testing CZT material and detectors that emphasize the particular roles of two types of extended defects, and their contributions to the device's overall performance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.