Abstract
To analyze the correlation between transcutaneous oxygen pressure (P(tc)O(2)) and blood lactate in patients with septic shock. Fifty-sixpatients with septic shock were prospectively investigated. P(tc)O(2) was monitored continuously for 6 hours, and arterial blood gas was measured at baseline (T0) and 6 hours(T6). Records of P(tc)O(2),were analyzed for the correlation with lactate level and lactate clearance rate. P(tc)O(2) valuesin the high lactate clearance group and the low one were compared.The lowest value of P(tc)O(2) at T6 and duration of P(tc)O(2)<40 mmHg (1 mmHg=0.133 kPa) were both correlated with lactate level and lactate clearance rateat T6.The low predictive value of P(tc)O(2) was 29 mmHg of lactate clearance under 20% with a sensitivity 85.2% and a specificity 65.5%. The low predictive value of P(tc)O(2) in high lactate clearance group was significantly higher than that in low lactate clearance group, while the duration of P(tc)O(2)<40 mmHg was shorter than the latter. During 6 h continuous monitoring, patients with a significant low P(tc)O(2) or prolonged duration of low P(tc)O(2) have relatively high lactate or low lactate clearance after resuscitation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.