Abstract

The purpose of this study was to measure staff radiation doses in representative institutions of five European countries performing coronary angiographies (CA) and percutaneous transluminal coronary angioplasties (PTCA), as well as to investigate any correlation with patient doses. Patient doses were recorded in terms of dose-area product (DAP). Staff doses were measured by thermoluminescence dosemeters placed on the operator's left shoulder and left foot. Radiation protection measures were also recorded. Third quartile shoulder and foot doses were 90 and 66 microSv in CA, and 157 and 173 microSv in PTCA, respectively. Shoulder dose was poorly correlated with DAP in CA (R2 = 0.29) and only moderately correlated in PTCA (R2 = 0.69). DAP and foot dose correlation was poor both in CA (R2 = 0.16) and in PTCA (R2 = 0.02). Protective measures taken by staff varied between countries, which may explain the poor correlation of patient and staff radiation doses.