Abstract

Although rate responsive pacing based on impedance minute ventilation (IMV) is now standard, there is almost no data confirming the relationship between IMV from an implanted pacemaker and measured minute ventilation (VE) during exercise. Nineteen completely paced adults implanted with Medtronic Kappa 400 pacemakers underwent symptom-limited maximal metabolic treadmill testing using a modified Minnesota Pacemaker Response Protocol. Minute ventilation (VE, L/min) was simultaneously measured using the flowmeter of a respiratory metabolic gas analysis system and the transthoracic impedance minute ventilation circuitry of the pacemaker. Correlation coefficients (r) were used to find the best fit line to describe the relationship between the two measurements. Mean (+/- SD) r values for the first, second, and third order polynomial equations and for log and exponential equations were: 0.92 +/- 0.08, 0.94 +/- 0.04, 0.95 +/- 0.04, 0.91 +/- 0.06, and 0.91 +/- 0.07, respectively. None of the r values were statistically different from the first order equation. Transthoracic IMV as measured by the Medtronic Kappa 400 is closely correlated to measured minute ventilation and is represented well by a first order (linear) equation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.