Abstract

Iridium oxide is routinely used for bionic applications owing to its high charge injection capacity. The electrode impedance at 1 kHz is typically reported to predict neural recording performance. In this article, the impedance of activated iridium oxide films (AIROFs) has been examined. The impedance of unactivated iridium electrodes was half that of platinum electrodes of similar geometry, indicating some iridium oxide was present on the electrode surface. A two time constant equivalent circuit was used to model the impedance of activated iridium. The impedance at low and intermediate frequencies decreased with increasing number of activation pulses and total activation charge. The impedance at 12 Hz correlated with the steady-state diffusion electroactive area. The impedance at 12 Hz also correlated with the charge density of the electrode. The high charge density and low impedance of AIROFs may provide improved neural stimulation and recording properties compared with typically used platinum electrodes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.