Abstract

Objective To investigate the influence of acquisition time of C-arm cone-beam CT on image quality and radiation dose of cerebral angiography.Methods C-arm cone-beam CT of cerebral angiography was performed on the male anthropomorphic head phantom,with DynaCT imaging mode and the acquisition time of 5 s,8 s and 20 s were used.Scanning was performed with each acquisition time for three times,and VR,MIP and MPR images were reconstructed.The attenuation values and their standard deviations of intracranial segment of the internal carotid artery (ICA),middle cerebral artery(MCA),anterior cerebral artery(ACA) and uniformed brain tissues were measured to calculate the image noise,signal to noise ratio (SNR) and contrast to noise ratio (CNR).We used the image noise,SNR and CNR as the objective standard to evaluate the image quality,and One-way ANOVA analysis of variance was used to assess the difference among them.A scale with scores 1 to 5 was used to rate the quality of the reconstructed image of ICA,MCA,ACA as a subjective evaluation,the difference among the evaluation scores were analyzed using Kruskal-wallis.We recorded the dose area product (DAP) of each acquisition time and the effective dose(ED) was calculated to assess the radiation dose,the difference among them were analyzed using One-way ANOVA analysis of variance.Results In C-arm cone beam CT 20 s imaging,the SNR of intracranial segment of the ICA,M CA,ACA were 22.29± 1.41,29.36 ±0.11 and 23.13 ±2.10 respectively,in 5 s imaging13.83 ±0.61,14.65 ±0.16 and 12.79±0.19 respectively,in 8 s imaging 14.92±0.96,18.97 ± 1.24 and 16.65 ±0.46 respectively,all the results showed a significant difference (F valued 58.19,327.29,52.74 respectively,all P valued<0.01),the CNR of 20 s imaging were higher than that of 5 s and 8 s imaging,the Noise of ICA,MCA,ACA and the uniformed brain tissues of 20 s imaging were lower than that of 5 s and 8 s imaging,all the results showed significant difference (all P valued<0.01).The subjective evaluation scores of VR imaging of 5 s,8 s and 20 s were 3.61 ±0.49,4.06±0.53,4.72±0.45 respectively,the scores of MIP imaging were 3.42±0.50,3.83±0.65 and 4.50±0.51 respectively,the scores of MPR imaging were 2.83±0.45,3.14±0.35 and 3.67±0.49 respectively,all the results showed significant difference (x2 valued 51.29,42.25 and 43.56 respectively,all P valued <0.01).The DAP values in C-arm cone beam CT 5 s,8 s and 20 s imaging were (9.11 ±0.18),(13.81±0.75)and(58.62±0.62)Gy· cm2,ED values were (0.91 ±0.02),(1.38±0.75)and(5.84± 0.11) mSv,all the results showed significant difference (F valued 3 720.30 and 3 654.85 respectively,all P valued<0.01).Conclusion The image quality and radiation dose of C-arm cone-beam computed tomography are closely correlated with acquisition time. Key words: Tomography,X-ray computed ; Angiography ; Image quality ; Radiation dose

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call