Abstract

Objective: The aim of this study was to determine the correlation between the distance from the file tip to the apical foramen and the readings on displays of two Electronic Apex Locators. Methodology: 46 human teeth were accessed, and electronic measurements were obtained at the display levels: 2.0; 1.5; 1.0; 0.5 and “apex”, these electronic measurements were repeated. Results: At the 0.0 and 0.5 display numbers, there was an accurate detection of the apical foramen: 93.5% for the Root ZX II and 93.5% for the RomiApex A-15 (± 0.5 mm). Analysis of variance (ANOVA) revealed that the accuracies of the two devices did not differ (p>0.05). At 2.0, 1.5 and 1.0, the accuracy was low and there was a significant difference in the accuracy of booth devices (p=0.000). When evaluating repetition of measurements, there was a high level of reproducibility (the ICCs ranged between 1.000 and 0.992 for the Root ZX II and between 1.000 and 0.989 for the RomiApex A-15) in all evaluated display’s levels. Conclusions: Both devices displayed high accuracy and reproducibility at the 0.0 and 0.5 display numbers.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.