Abstract
By investigating the correlation between the injection rate and pressure of subretinal tissue plasminogen activator (tPA) and air using a standard Viscous Fluid Control (VFC) system with a 38-gauge cannula, we aimed to establish guidelines for stable injections. We fabricated a retina mimicking model (RMM) with 0.25% agarose solution and an aluminum plate, and substituted submacular hemorrhage (SMH) and tPA with blood-mimicking fluid (BMF) and balanced salt solution (BSS), respectively. The diameter of the pre-bleb mimicking SMH in RMM was 1.30 ± 0.16 cm, increasing to 1.98 ± 0.24 cm and 1.83 ± 0.22 cm after bleb propagation with BSS and air, respectively. BSS injection rates were 2.86 ± 0.04 µl/sec, 6.74 ± 0.48 µl/sec and 8.55 ± 0.16 µl/sec at 8, 12, and 16 psi, respectively. Air injection rates were 37.98 ± 3.11 µl/sec, 79.01 ± 5.13 µl/sec and 156.06 ± 13.72 µl/sec at 2, 3 and 4 psi, respectively. By experimenting with different pressures in the RMM, we found 12 psi to be the minimum for proper BSS injection and 2 psi for air. These findings provide crucial parameters for safer surgery to prevent irreversible damage.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.