Abstract

We sought to assess the accuracy of a novel parameter proportional to the rod shear stress (RSS) in identifying patients at risk of rod fracture (RF) after surgery for correction of adult spinal deformity. We performed a retrospective medical record review of patients aged ≥18 years treated for adult spinal deformity between 2004 and 2014 with ≥24 months of follow-up. The primary outcome was RFs identified radiographically. Patient weight (w), number of instrumented levels (N), and minimum rod diameter (d) were recorded and used to calculate the RSS parameter (RSS=Nwd2). Receiver operating characteristic curves were produced and the area under the curve (AUC ± 95% confidence interval [CI]) was calculated to compare this parameter's discriminative accuracy to that of its constituent variables. The sensitivity, specificity, and likelihood ratios (LRs) were calculated. A total of 28 RF-positive and 154 RF-negative patients were included. The average age was 59.2 ± 9.6 years, and 93.4% were women. The RSS parameter produced the greatest AUC (0.73 ± 0.11). At an RSS cutoff of 30.1, it achieved a sensitivity of 71.4% and specificity of 71.4% (LR, 2.5; 95% CI, 1.8-3.5). The number of instrumented levels produced the next-greatest AUC (0.65 ± 0.12), with a sensitivity of 78.6% and specificity of 50.0% at a cutoff of 15 (LR, 1.6; 95% CI, 1.2-2.0). The RSS is calculated using easily obtainable information and shows potential as a tool for predicting patient-specific risk of RF after spinal fusion. The number of instrumented levels also correlates strongly with the occurrence of RFs and is not significantly less accurate than the RSS. A larger sample size and prospective validation would be useful in determining with greater confidence which parameter is superior for predicting RFs after spinal fusion.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call