Abstract

To determine the correlation between the primary implant stability quotient and the implant percussion sound frequency. A total of 14 pigs' ribs were scanned using a dental cone beam computed tomography (CBCT) scanner to classify the bone specimens into three distinct bone density Hounsfield units (HU) value categories: D1 bone: >1250 HU; D2: 850-1250 HU; D3: <850 HU. Then, 96 implants were inserted: 32 implants in D1 bone, 32 implants in D2 bone, and 32 implants in D3 bone. The primary implant stability quotient (ISQ) was analyzed, and percussion sound was recorded usinga wireless microphone connected and analyzed with frequency analysis software. Statistically significant positive correlations were found between the primary ISQ and the bone density HU value (r = 0.719; p < 0.001), and statistically significant positive correlations between the primary ISQ and the percussion sound frequency (r = 0.606; p < 0.001). Furthermore, significant differences inprimary ISQ values and percussion sound frequency were found between D1 and D2 bone, as well as between D1 and D3 bone. However, no significant differences were found in primary ISQ values and percussion sound frequency between D2 and D3 bone. The primary ISQ value and the percussion sound frequency are positively correlated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.