Abstract

Head motion during Computer Tomographic (CT) studies can adversely affects the reconstructed image through distortion and other artifacts such as blurring and doubling, thereby contributing to misdiagnosis of diseases. In this paper, we propose a method to detect and mitigate motion artifacts in three-dimensional (3D) cone-beam CT system. Motion detection is achieved by comparing the correlation coefficient between the adjacent x-ray projections. Artifacts, caused by motion, are mitigated either by replacing motion corrupted projections with their counterpart 180° apart projections under certain conditions, or by estimating motion corrupted projections using Least Square Support Vector Machine (LS-SVM) based time series prediction. The method has been evaluated on 3D Shepp-Logan phantom. In this research, Feldkamp-David-Kress (FDK) based back-projection algorithm is used for 3D reconstruction process. Computer simulation validates the motion detection and artifacts elimination mechanism.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.