Abstract

Neutron activation analysis using the single comparator method or the k 0 method may be less accurate in the cases of non-1/v nuclides if irradiations are done at varying reactor temperatures. We propose a method to correct this effect, using available reactor moderator temperature readings. It was verified by the irradiation of two non-1/v nuclides, 176Lu and 151Eu. Irradiations were carried out in two irradiation sites at two reactor powers and at estimated neutron temperatures varying from 18 to 46 °C. In all cases the measured activities showed a variation with temperature consistent with published tables of the variation of the Westcott g-factor. The method should be accurate to 1% for most small reactors and eliminates the need to irradiate temperature monitors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.