Abstract

Single molecule photobleaching is a powerful technique to measure the number of fluorescent units in subresolution molecular complexes, such as in toxic protein oligomers associated with amyloid diseases. However, photobleaching can occur before the sample is appropriately placed and focused. Such “prebleaching” can introduce a strong systematic bias toward smaller oligomers. Quantitative correction of prebleaching is known to be an ill-posed problem, limiting the utility of the technique. Here, we provide an experimental solution to improve its reliability. We chemically construct multimeric standards to estimate the prebleaching probability, B. We show that B can be used as a constraint to reliably correct the statistics obtained from a known distribution of standard oligomers. Finally, we apply this method to the data obtained from a heterogeneous oligomeric solution of human islet amyloid polypeptide. Our results show that photobleaching can critically skew the estimation of oligomeric distributions, so that low abundance monomers display a much higher apparent abundance. In summary, any inference from photobleaching experiments with B > 0.1 is likely to be unreliable, but our method can be used to quantitatively correct possible errors.