Abstract
Background Rabbits are commonly used as animal models for the evaluation of drugs and surgery to lower intraocular pressure (IOP).The accuracy of IOP measurement is therefore critical in the analysis of data and subsequent extrapolation to humans.An accurate method to measure rabbit IOP is intracameral manometry,but it is an invasive way.Schiotz,Perkins and Rebound were often used in clinic.However,their accuracy in measuring rabbit IOP in experimental study is unclear. Objective The purpose of this study was to investigate the accuracy of IOP measured by Schiotz tonometer,Perkins tonometer and Rebound tonometer relative to intracameral manometry in New Zealand white rabbits. Methods The central corneal thickness (CCT),corneal curvature (CC),axial length (AL),anterior chamber depth (ACD), lens thickness (LT) and scleral thickness (ST) were respectively measured in 8 eyes of 8 healthy New Zealand white rabbits with lenstar900 and ultrasound biomicroscopy (UBM).The actual IOP was measured with a 24G needle inserted the anterior chamber and connected to a pressure transducer under the general anesthesia,the IOP gradient was set with a 24G needle inserting the vitreous cavity and connecting to a container with balanced salt solution(BSS).Then,comparative measurements at the same pressures were performed with three types of tonometers.The IOP values from Schiotz tonometer,Perkins tonometer and Rebound tonometer were calibrated based on actual IOP from intracameral manometry and eyeball physiological parameters by multiple regression equation. Results The mean of CCT,CC,AL,ACD,LT and ST was (338.96±21.52) μm,(51.68±1.66) D,(14.63±0.19) mm,(2.22±0.04) mm,(6.15±0.10) mm and (339.80±47.41) μm.Compared with the intracameral manometry value (IMV),the error range was (17.08±11.22) mmHg in the Schiotz tonometer value (STV),(25.81±12.43) mmHg in the Perkins tonometer value (PTV) and (22.50±11.47) mmHg in the Rebound tonometer value (RTV),with significant differences between them(t=10.54,14.39,13.59,all at P<0.05).Compared with IMV,the 95% limits of agreement of three portable tonometer values is larger,and three portable tonometer values had the greater measurement error with elevated IOP gradient.The regression equations was IOP=141.015+1.570×STV+0.122×CCT-3.480×CC between actual IOP and STV (R=0.92,P=0.00),IOP=-33.323+1.914×PTV+0.133×CCT between actual IOP and PTV(R=0.88,P=0.00),IOP=160.395+1.866×RTV+0.201×CCT+34.554×LT-2.649×CC+0.063×ST between actual IOP and RTV (R=0.95,P=0.00). Conclusions The physiological parameters of rabbit eyeball are obviously different from human. The STV, PTV and RTV have a great measuring error in comparson with actual IOP, and therefore it is necessary to correct STV, PTV and RTV based on the ocular physiology parameters in experimental study. Key words: Bio-measurment; Intraocular pressure; Intracameral manometry; Tonometer; Regression equation; Rabbit
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.