Abstract

Fluorescence excitation-emission matrix (EEM) spectroscopy is a useful tool for interpretation of fluorescence information from natural water samples. One of the major problems with this technique is the inner-filter effect (IFE), i.e. absorption of light at both the excitation and emission wavelengths. The common solutions are to either dilute the sample or apply some form of mathematical correction, most often based on the measured absorbance of the sample. Since dilution is not always possible, e.g . in on-line or in situ EEM recordings, and corrections based on absorbance are hampered primarily by the use of a separate absorbance instrument, neither of these solutions is optimal. In this work, we propose a mathematical correction procedure based on the intensity of Raman scatter from water. This procedure was found to reduce the error after correction by up to 50% in comparison with two absorbance correction procedures. Furthermore, it does not require the use of a separate absorbance measurement, and it is applicable to on-line and in situ EEM recordings, where the IFE would otherwise cause problems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.