Abstract
With the help of a fundamental parameter program, the matrix effect of multi-component complex samples and the composition of each element’s X-ray fluorescence intensity were calculated. A mathematical model for correcting the multi-component matrix effect was deduced theoretically. It provides two methods for calculating theoretical influence coefficients α and β. The correction model and the influence coefficient introduced in this paper have clear and definite physical meanings. They gave satisfactory results when utilized in analyzing stainless-steel samples. The results also indicate that the effect of tertiary fluorescence cannot be neglected in some cases. From the results of comparison experiments, it is concluded that the tertiary fluorescence effect must be taken into consideration while analyzing Cr in a steel sample with a high concentration of Ni. © 1998 John Wiley & Sons, Ltd.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.