Abstract

The response of electronic portal imaging devices (EPIDs) of the scanning liquid ionization chamber (SLIC) type is known to vary with linear accelerator gantry angle. This work considered several contributing factors, quantified the artefacts, monitored their reproducibility and investigated the effects of repeated gantry rotations. Unflatness of up to 5% was found. A correction technique was devised using nonlinear regression of a three-variable sinusoidal modulation. Comparison with two existing techniques found our method to be the most effective, providing a flatness well within 2%. This improved accuracy is expected to benefit more accurate dosimetric studies in particular. The post-acquisition correction process required no change in imaging protocols. Applicability of the new technique was demonstrated on images acquired on different days and with different beam sizes. Since the artefacts compromise both accurate dosimetry and image quality, their successful removal should benefit a broad range of SLIC EPID applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.