Abstract

A basic result in gamma spectrometry is the count rate of a relevant peak. Correction for decay during counting and expressing the count rate at the beginning of the measurement can be done by a multiplicative factor that is derived from integrating the count rate over time. The counting time substituted in this factor must be the live time, whereas the use of the real-time is an error that underestimates the count rate by about the dead-time (DT) (in percentage). This error of underestimation of the count rate is corroborated in the measurement of a nuclide with a high DT. The present methodology is not applicable in systems that include a zero DT correction function.