Abstract
Analytical results are presented obtained with Zeeman atomic absorption (ZAA) using a modified sine wave magnetic field. The measurement at zero field strength is lengthened to 0.5 ms for a 50 Hz magnetic field of 10 kG. The a.c. Zeeman system is extended with an additional intensity measurement performed at an intermediate field strength. The three field Zeeman system (3FZAA) permits simultaneous correction for background absorption and stray radiation at the expense of halved analytical sensitivity. The background correction capabilities of ZAA and 3FZAA are the same. However, the 3FZAA signals show increased noise in comparison to ZAA. In the three field system the roll-over problem, inherent in existing Zeeman systems, is shifted to higher concentration and to higher absorbance. The height and the position of the maximum in ZAA and 3FZAA analytical curves do not depend on the amount of background absorption. A method for the extension of analytical curves in AA is presented. Utilizing an a.c. modulated magnetic field any sensitivity between zero and ordinary AA sensitivity can be obtained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.