Abstract
Abstract Context.—Correction factors have been proposed for estimating true potassium concentrations in blood samples with evidence of in vitro hemolysis. Objective.—We used 2 different models of true (ie, nonsimulated) in vitro hemolysis to evaluate the clinical utility of correction factors for estimating potassium concentrations in samples with evidence of in vitro hemolysis. Design.—Potassium correction factors were derived using 2 different models. In model 1, potassium and plasma hemoglobin were measured with the Hitachi 747 analyzer in 132 paired blood samples, with each pair consisting of 1 sample with evidence of hemolysis and 1 without, collected during the same phlebotomy procedure. The change in measured potassium concentration was plotted versus the change in plasma hemoglobin concentration for each pair of samples. In model 2, the potassium levels of 142 784 blood samples and the corresponding hemolytic index values were measured with the Beckman LX20 analyzer. Potassium concentrations at th...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.