In the past two decades, the water calorimetry technique for determination of absorbed dose to water in several types of radiation beams has moved significantly closer to being a recognized method. In this paper we summarize the constructional details of a C sealed water calorimeter currently in operation at the University of Gent. This sealed water (SW) calorimeter is of the Domen type and has been improved in several aspects compared with its original design. The relevant correction factors for heat transport and for field perturbation are described. Using relative response measurements in , we experimentally verified the relative heat defect for two distinct chemical systems, using two different detection vessel arrangements. The overall uncertainty on the absorbed dose to water at based on this system amounts to 0.7%.