Abstract

The finite number of pixels in a silicon photomultiplier (SiPM) limits its dynamic range to light pulses up to typically 80% of the total number of pixels in a device. Correcting the non-linear response is essential to extend the SiPM's dynamic range. One challenge in determining the non-linear response correction is providing a reference linear light source. Instead, the single-step method used to calibrate PMTs is applied, based on the difference in responses to two light sources. With this method, the response of an HPK SiPM (S14160-1315PS) is corrected to linearity within 5% while extending the linear dynamic range by a factor larger than ten. The study shows that the response function does not vary by more than 5% for a variation in the operating voltage between 2 and 5 V overvoltage in the gate length between 20 and 100 ns and for a time delay between the primary and secondary light of up to 40 ns.