Abstract

The accurate assessment of visual field function can provide valuable information on a range of visual disorders. Saccadic Vector Optokinetic Perimetry (SVOP) is a novel instrument for measuring supra-threshold visual fields in young children who are otherwise unable to perform Automated Static Perimetry (ASP). However, limitations in Liquid Crystal Display (LCD) technology restrict the ability of SVOP to determine threshold values at various points in the visual field, often required in detailed perimetry examinations. This paper introduces a purpose-specific LCD luminance non-uniformity compensation approach to address this limitation. Thorough quantitative evaluation identifies the effectiveness of the proposed approach in (i) compensating for luminance non-uniformities across an LCD, and (ii) enabling SVOP to perform accurate and precise threshold visual field tests. The findings demonstrate that SVOP provides a promising alternative to the current threshold ASP standard (Humphrey Field Analyser).