Abstract
The original description of the TIMI frame count (TFC) method was based on angiograms acquired at 30 f/sec. Modern digital angiograms are acquired at lower frame rates (between 12.5 and 25 f/sec). Coronary angiography was acquired at 12.5 and 25 f/sec after 200 mug of intracoronary glyceryl trinitrate. Results of the corrected TIMI frame count (cTFC) at 12.5 and 25 f/sec for each vessel were: right coronary artery, 19.5 +/- 5.2 and 20.4 +/- 6.6 (P = 0.15); circumflex artery, 25.6 +/- 8.2 and 25.9 +/- 8.7 (P = 0.5); and left anterior descending artery, 22.5 +/- 8.1 and 23.8 +/- 10.4 (P = 0.15), respectively. The mean difference in the TFC between two injections by the same operator and by two operators was 0.4 (P = 0.7) and 0.4 (P = 0.2), respectively. The mean difference in the TFC for repeat measurements by the same observer and between two observers was 0.26 (P = 0.3) and 0.06 (P = 0.8), respectively. We confirm that the cTFC is a quantitative method to assess coronary flow that can be applied in a modern digital laboratory.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.