Abstract

Coronary stent fracture (SF) is a potential cause of stent failure increasing the risk for in-stent restenosis, stent thrombosis, target lesion revascularization and major adverse cardiac events. Overall incidence of SF ranges from 1.0% up to 18.6% and SF can be found in up to 60% of failed devices. Advanced imaging techniques have improved the detection of SF. However, defining the optimal therapeutic approach towards these complex lesions is challenging. This review summarizes the most important publications on the topic of SF and discusses current insights into pathophysiology, diagnostic tools, classification and therapeutic management. Furthermore, two illustrative cases of SF leading to myocardial infarction are presented, which demonstrate typical SF risk factors such as vessel angulation and hinge motion, stenting in the right coronary artery, use of long stents and multiple stent layers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.