Abstract

We report a series of three patients whose implantable cardioverter-defibrillators (ICD) implants were unsuccessful due to inability to achieve defibrillation thresholds (DFT) at maximum available energy after failure of standard modification and enhancement procedures. All patients had advanced cardiomyopathy. Use of the coronary sinus (CS) for left ventricular (LV) shocking electrode placement resulted in acceptable DFTs in each patient. The position of the shocking coil in all three patients was posterior, and in two patients alongside a left ventricular CS pacing lead. The best shocking configuration tested was LV (CS) + CAN (Anode) to RV (cathode) in each patient. The short- and long-term outcomes of these patients is presented and discussed. This approach is suggested as a salvage option for those problematic patients who have unacceptable DFT results at implantation of an endovascular ICD system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.