Abstract

The inability to return the heart to the same position for all breath-holds during 2D coronary MR imaging can result in imaging different locations than desired. This can lead to problems such as (i) missing a whole vessel, or a part of it, (ii) misaligning segments of vessels imaged in different breath-holds, and (iii) degrading image quality when a single slice is acquired in multiple breath-holds. To reduce inconsistencies in the breath-hold level, we designed a respiratory feedback monitor (RFM) that uses a bellows to monitor the circumference of the subject's chest. When the circumference of the subject's chest is within preset limits, an audio signal alerts subjects to hold their breath at that position. Use of the RFM significantly reduces the problems caused by inconsistent breath-holds and the number of breath-holds for an examination in 2D coronary MR imaging.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.