Abstract

Support catheters with a threaded, distal tip may be useful for crossing recalcitrant coronary lesions. The interaction between the Gopher support catheter and polymer-coated coronary guidewires may, however, result in circumferential stripping of polymer from the guidewire, with resulting wire damage or fracture. We report two cases of polymer-coated guidewire damage with use of a Gopher support catheter, suggesting that these two devices should not be used together.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.