Abstract

The aim of this study was to compare the calcium score and reproducibility of coronary artery calcium scores obtained on the four kinds of 64-detector computed tomography (CT) scanners using standard (120 kVp) and low tube voltage (80 kVp) scan techniques. We scanned 80 and 120 kVp on all scanners. We calculated Agatston, volume, and mass scores for coronary artery calcium scoring on each scanner and compared the coefficients of variation of the calcium scores to evaluate reproducibility of among CT scanners. The averages of the total mean Agatston score, total mean volume score, and total mean mass score at 80 kVp/120 kVp were 798.9/683.8, and 627.2/567.3, and 157.1/156.7, respectively. The total mean mass score was almost constant irrespective of the tube voltage. The total mean coefficients of variation for the four CT scanners were lower at 80 than 120 kVp (4.1% vs. 10.2% [total mean Agatston score], 3.2% vs. 9.6% [total mean volume score], and 3.2% vs. 9.4% [total mean mass score]). Use of the low tube voltage technique can reduce variations in the coronary artery calcium scores obtained on different CT scanners.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.